Maintenance
DVVA II 4000
9-20
400-42803-01 (Rev B) - July 2017
10. Slide the sample compartment shield up; wipe the piston tip with a Kimwipe or paper towel.
11. Slide the sample compartment shield down; press
Reset
.
12. Select
Unit 1 > Sample Analysis
; the Analysis dialog is displayed with the Open Sample
Information dialog positioned on top.
13. Click
OK
to accept the next sequenced number in the
File name
field; the Analysis dialog
containing file defaults is displayed.
14. Click
Replace All
and select the file named
COND.smp
, then click
OK
.
All appropriate values will be inserted into the dialog fields.
15. Edit the description in the
Sample field
by adding a
1
indicating this is the first of 10
analyses; 10 analyses are performed when conditioning the piston tip.