DVVA II 4000
Appendix A
400-42803-01 (Rev B) - July 2017
A-17
6007-
No data available in file (file name) for reporting.
6017- Lower
piston motor failure.
6018-
The lower piston cannot be moved because it is under load. Relieve the pressure
on the lower piston and try again.
6019- The lower piston cannot be moved because the upper piston is too low. Raise the
upper piston into the chamber and try again.
6020-
The upper piston cannot be moved down with the lower piston in its current posi-
tion. The upper piston will only move down when the lower piston is either fully
open or fully closed.
Cause:
You requested a report for a file that does not contain collected data.
Action:
Choose a file that contains collected data (
Complete
status).
Cause:
The lower piston cannot drive to the requested position.
Action:
Contact your service representative.
Cause:
You attempted to move the lower piston but its load is too high.
Action:
Raise the upper piston so that it is not pressing against the lower one.
Cause:
The lower piston will not move because the upper piston tip is too low.
Action:
Raise the upper piston above 0.025 inch to allow the lower piston to
move.
Cause:
You attempted to move the upper piston when the lower piston was not
in the correct position.
Action:
Open or close the lower piston completely. The lower piston must be
fully open or completely closed for the piston to move.
Содержание DVVA II 4000
Страница 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Страница 4: ......
Страница 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Страница 16: ......
Страница 32: ......
Страница 116: ......
Страница 136: ......
Страница 160: ......
Страница 180: ......
Страница 212: ......
Страница 214: ......
Страница 240: ......
Страница 246: ......
Страница 248: ......