DVVA II 4000
Appendix A
400-42803-01 (Rev B) - July 2017
A-7
2482- File cannot be opened for writing. It is already in use.
2483- An analysis cannot be performed on (file name). It is open for editing and contains
errors.
2484-
The edit session for (file name) must be saved before the analysis. Save changes
and continue with the analysis.
Cause B:
A computer or network problem occurred.
Action B:
Check the performance of your computer devices or network.
Cause C:
A software error occurred when the file was accessed.
Action C:
Contact your service representative.
Cause:
You are attempting to open a file that is currently being used (either by
this application or another).
Action:
Locate the application using the file (in the Micromeritics application,
use the Windows menu item to get a list of all windows, one of which
may contain this file).
Cause:
You attempted to use a sample file containing errors that is currently
open.
Action:
Go to the window containing the file, correct the errors, and save it.
Cause:
You attempted an analysis using a sample file that contains unsaved
changes.
Action:
Select
Yes
to save the changes and continue with the analysis.
Select
No
to abort the analysis and return to the sample file.
Содержание DVVA II 4000
Страница 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Страница 4: ......
Страница 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Страница 16: ......
Страница 32: ......
Страница 116: ......
Страница 136: ......
Страница 160: ......
Страница 180: ......
Страница 212: ......
Страница 214: ......
Страница 240: ......
Страница 246: ......
Страница 248: ......