Appendix A
DVVA II 4000
A-14
400-42803-01 (Rev B) - July 2017
2534-
Error opening file (name) for printing. Error code: (number).
2547- The instrument is busy. Exiting the application will result in the termination of the
analysis. Continue with program exit? (Yes, No)
2549- Error accessing online manual file (name).
Cause:
An error occurred in the selected file for print output.
Action:
Ensure that sufficient space is available on the drive containing the file.
Cause:
You attempted to exit the application while an analysis is in progress.
Action:
If you are not concerned with cancellation of the analysis, click
Yes
to
continue the exit; otherwise select
No
and allow the analysis to finish.
Cause:
You were unable to open the online manual.
Action:
Exit the application and reinstall the software. The manual and/or
reader may have been accidentally deleted. Reinstalling the software
will reinstall the manual and reader.