Setting up a MRM Function
Multiple reaction monitoring (MRM) functions are set up in much the same way as
SIR functions, but allow a number of MS-MS transitions (fragmentations) between
MS1 and MS2 to be monitored.
All fields in the MRM editor are similar to those already described.
Setting up a Survey Function
Survey scans are used to search for precursor ions. To access the dialog:
Press
or select
Survey Scan
from the
Functions
menu in the
scan functions editor.
The function list editor does not add survey functions to the list if non-survey
functions are present.
Data Acquisition
Page 93
Quattro Ultima
User's Guide