Disconnect both electrical connections by pulling back on the plug sleeves to
release the plugs from the sockets on the front panel.
Undo the two knurled thumb nuts that retain the probe and withdraw it from the
source. Place it carefully to one side.
Undo the three thumb screws and withdraw the probe adjustment flange and
glass tube. Place the glass tube, end on, on a flat surface and place the probe
adjustment flange on top of the glass tube.
Warning:
When the source enclosure has been removed the source block is
exposed. Ensure that the source block heater has cooled before proceeding.
If fitted, remove the APcI discharge pin.
The sample cone and cone gas nozzle are now accessible.
Using a suitable flat blade screwdriver rotate the isolation valve by 90° into its
fully anticlockwise position.
A small improvement in the analyser vacuum may be observed as a result of this
operation.
The isolation valve is closed when the slot is perpendicular to the direction of
flow.
Maintenance and Fault Finding
Page 185
Quattro Ultima
User's Guide
Cone Gas
Nozzle
Isolation
Valve
Содержание Quattro Ultima
Страница 2: ...Quattro Ultima User s Guide...
Страница 68: ......
Страница 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Страница 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Страница 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Страница 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Страница 222: ...Reference Information Page 222 Quattro Ultima User s Guide...