Manual Editing of Peak Matching
If an incorrect peak has been matched in the calibration process, this peak can be
excluded manually from within the on-screen calibration report.
Using the mouse place the cursor over the peak in the acquired spectrum and
click with the right mouse button.
The peak is excluded and is no longer highlighted.
If the true reference peak is present then this can be included in the calibration by the
same procedure.
Place the cursor over the required peak and click with the right mouse button.
The peak is matched with the closest peak in the reference spectrum.
Manually editing one peak does not affect the other matched peaks in the calibration.
Saving the Calibration
When the instrument is fully calibrated the calibration can be saved under a file name
so that it can be recalled for future use.
The recalled calibration has the same constraints of mass range and scan speed. The
ion energy and resolution settings used for the calibration acquisition are also recorded
as these can have an effect on mass assignment.
Mass Calibration
Page 121
Quattro Ultima
User's Guide
Содержание Quattro Ultima
Страница 2: ...Quattro Ultima User s Guide...
Страница 68: ......
Страница 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Страница 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Страница 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Страница 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Страница 222: ...Reference Information Page 222 Quattro Ultima User s Guide...