Preparation for APcI Operation
If the corona discharge pin is not fitted, proceed as follows:
Disconnect the gas and electrical connections from the front panel.
Unscrew the probe thumb nuts and remove the probe.
Undo the three thumb screws and remove the probe adjustment flange and
glass tube.
Remove the blanking plug from the discharge pin mounting contact and fit
the corona discharge pin, ensuring that the tip is in-line with the tip of the
sample cone.Connect the APcI high voltage cable between
Capillary / Corona
and the socket positioned at the lower right edge of
the source flange.
Replace the glass tube and adjustment flange.
Ensure that the source enclosure (consisting of the glass tube and the probe
adjustment flange) is in place.
Connect the source’s gas line to
Desolvation Gas
on the front panel. Tighten
the nut to ensure a good seal.
Check that the lead of the probe adjustment flange is plugged into the socket
labelled
ESI / APcI
on the front panel.
Routine Procedures
Page 40
Quattro Ultima
User's Guide
Source
Thumb Nuts
Probe
Thumb Nuts
Probe
Adjustment Flange
Glass
Tube
Содержание Quattro Ultima
Страница 2: ...Quattro Ultima User s Guide...
Страница 68: ......
Страница 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Страница 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Страница 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Страница 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Страница 222: ...Reference Information Page 222 Quattro Ultima User s Guide...