Calibration Options
To access the calibration options:
Select
Calibration
then
Calibrate Instrument...
on the tune page.
Selecting the Reference File
To select the appropriate reference file:
Click on the arrow in the
Reference File
box and scroll through the files.
Select
nairb.ref
for a sodium iodide and rubidium iodide reference solution.
Removing Current Calibrations
Select the
uncal.cal
calibration file from the
File
,
Load Calibration...
menu
option.
Select
Process
,
Delete all calibration...
followed by
File
,
Save Calibration
.
This ensures that a file with no calibration is currently active on the instrument
and prevents any previously saved calibrations from being modified or
overwritten.
Mass Calibration
Page 106