Megaflow
Megaflow electrospray enables flow rates from 200 µl/min to 1 ml/min to be
accommodated. This allows microbore (2.1mm) or 4.6mm diameter columns to be
interfaced without splitting.
Changing Between Flow Modes
When changing between megaflow and standard electrospray operation, it is essential
that the correct tubing is used to connect the probe to the sample injector. For
megaflow operation
1
/
16
“ o.d., 0.007" i.d. peek tubing, easily identified by its yellow
stripe, is used. This replaces the standard fused silica tube, together with the PTFE
sleeves.
Electrospray
Page 145
Quattro Ultima
User's Guide
Probe
PTFE Sleeve
PTFE Sleeve
Fused Silica Tube
1/16" o.d. 0.007" i.d. Peek Tube
Injector
Normal Flow Electrospray
Megaflow Electrospray