Remove the two screws from the heater connections on the ion block.
Carefully straighten the heater supply leads, in such a way that the ion block can
later be withdrawn without fouling these leads.
Loosen the screw on the thermocouple’s securing clip and unhook the
thermocouple from its location.
Remove the two screws which secure the ion block to the ion block support.
Withdraw the ion block, leaving the thermocouple and heater supply leads
protruding from the ion block support.
Ensure that the three O rings remain in position on the ion block support.
Maintenance and Fault Finding
Page 190
Quattro Ultima
User's Guide
Ion Block
Support
O Rings
O Ring
Restrictor
and Bush
Ion Block
Thermocouple
Heater
Leads
Heater Supply
Leads
PTFE
Washer