Set
Capillary
between 1 and 1.5kV.
Adjust
Desolvation Gas
using the knob on the front panel of the instrument.
An ion beam should now be visible on the tune page.
Tune the source voltages, adjust the gas flow and adjust the three-axis
manipulator for maximum ion current.
The ion current may change dramatically with very slight changes of position
but the high resolution of the threads in the manipulator allows very fine tuning.
Restarting the Spray
Should the spray stop, it is possible to restart it by adjusting the three-axis manipulator
so that, viewed under magnification, the capillary tip touches the sample cone and a
small piece of the glass hair shears off.
It may also be necessary to apply some back pressure to the holder to force a drop of
liquid from the capillary. Up to 1.4 bar (20 psi) can be applied and, with this pressure,
a drop should be visible unless the capillary is blocked.
Nanoflow Electrospray
Page 165
Quattro Ultima
User's Guide