![Micromass Quattro Ultima Скачать руководство пользователя страница 220](http://html1.mh-extra.com/html/micromass/quattro-ultima/quattro-ultima_user-manual_1787371220.webp)
Preparation of Calibration Solutions
PEG + Ammonium Acetate for Positive Ion Electrospray and APcI
Prepare a solution of polyethylene glycols at the following concentrations:
PEG 200
25 ng/µl
PEG 400
50 ng/µl
PEG 600
75 ng/µl
PEG 1000
250 ng/µl
Use 50% acetonitrile and 50% water containing 2 mmol ammonium nitrate.
Use reference file PEGH1000.
PEG + Ammonium Acetate for Positive Ion Electrospray
(Extended Mass Range)
Prepare a solution of polyethylene glycols at the following concentrations:
PEG 200
25 ng/µl
PEG 400
50 ng/µl
PEG 600
75 ng/µl
PEG 1000
250 ng/µl
PEG 1450
250 ng/µl
Use 50% acetonitrile and 50% water containing 2 mmol ammonium nitrate.
Use reference file PEGH2000