For APcI operation, the desolvation gas is not heated in the desolvation nozzle.
However, it is important that desolvation gas is used throughout.
The background spectrum for 50:50 acetonitrile:water is dependent upon the setting of
Cone
. The main reagent ions for a typical sample voltage of 40V are 83, 101 and
142.
Acetonitrile adducting may be minimised by optimisation of the cone gas and RF Lens
1 voltage, as described in
Electrospray
.
Preparation
Ensure that the source is assembled as described in
Maintenance and Fault
Finding
, and that the instrument is pumped down and prepared for APcI
operation as described in
Routine Procedures
.
APcI may be operated with or without the cleanable baffle fitted.
Ensure that a supply of nitrogen has been connected to the gas inlet at the rear of
the instrument and that the head pressure is between 6 and 7 bar (90-100 psi).
Atmospheric Pressure Chemical Ionisation
Page 170
Quattro Ultima
User's Guide
Corona
Discharge
Pin
Mounting
Contact
Sample
Cone
High Voltage
Socket
Blanking
Plug
Содержание Quattro Ultima
Страница 2: ...Quattro Ultima User s Guide...
Страница 68: ......
Страница 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Страница 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Страница 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Страница 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Страница 222: ...Reference Information Page 222 Quattro Ultima User s Guide...