It is wise periodically to check the flow rate from the solvent delivery system. This
can be carried out by filling a syringe barrel or a graduated glass capillary with the
liquid emerging from the probe tip and timing a known volume, say 10µl. Once the
rate has been measured and set, a note should be made of the back pressure readout on
the pump as fluctuation of this reading can indicate problems with the solvent flow.
Electrospray
Page 143
Quattro Ultima
User's Guide