Mass Measure Parameters
These are accessed through
Edit
,
Mass Measure Parameters...
. If
continuum or MCA data are acquired
for calibration then these parameters
need to be set before the calibration is
carried out. If centroided data are used
for calibration then the mass measure
parameters are not used.
With electrospray calibrations,
particularly with sodium iodide which
has some low intensity peaks at higher
mass, it is recommended that continuum
or MCA data are acquired.
At high scan speeds instrument
resolution may decrease. Ensure that the
centroiding parameters are set to use the
top 80% (or lower, if appropriate) of
the peak. Use
Centoid top (%)
, not
Top.
Mass Calibration
Page 109
Quattro Ultima
User's Guide