A calibration curve is produced and displayed on the screen in a similar way to when
the original calibration was performed. An example is shown above. When
OK
is
selected from this report, unlike the original calibration procedure, the instrument
calibration is not changed. As the verification procedure uses the same matching
parameters as the calibration procedure, it is possible to validate the current calibration
without re-calibrating the instrument.
The report can be printed out by selecting
,
from the verify report.
Mass Calibration
Page 139
Quattro Ultima
User's Guide