Select
Gas
,
Gas
to turn on the nitrogen.
Raise
APcI Probe Temp
gradually, starting at 100°C and increasing in 50°C
intervals to 650°C over a period of 10 minutes.
Caution:
Do not set
APcI Probe Temp
to 650°C immediately as this may
damage the probe heater.
This procedure should remove any chemical contamination from the probe tip.
Replacing the Probe Tip Heater
Remove the probe tip assembly by carefully loosening the two grub screws.
Disconnect the heater from the probe body by pulling parallel to the axis of the
probe.
Fit a new heater assembly.
Reconnect the probe tip assembly.
Maintenance and Fault Finding
Page 201
Quattro Ultima
User's Guide
Grub
Screws
Heater
Probe Tip
Assembly