Flow Control Valves
The
Desolvation Gas
,
Cone Gas
and
Nebuliser
needle valves are five-turn
valves. The
CID Gas
valve is a fifteen-turn valve. The flow increases as the valve is
turned counterclockwise.
Caution:
To prevent damage to the
CID Gas
valve, take care not to
over-tighten when turning the supply off.
Divert / Injection Valve
The optional divert / injection valve may be used in several ways depending on the
plumbing arrangement:
•
As an injection valve, with the needle port and sample loop fitted.
•
As a divert valve, to switch the flow of solvent during a LC run.
•
As a switching valve to switch, for example, between a LC system and a syringe
pump containing calibrant.
This valve is pneumatically operated, using the instrument’s nitrogen supply.
Note that the valve is connected such that the nitrogen supply is always
connected to the valve, irrespective of the flow to the source and probe.
Control of the valve is primarily from the data system. The two switches marked
Load
and
Inject
enable the user to override control of the valve when making loop
injections at the instrument.
Instrument Description
Page 25
Quattro Ultima
User's Guide
Divert / Injection
Valve
Load
Inject
Содержание Quattro Ultima
Страница 2: ...Quattro Ultima User s Guide...
Страница 68: ......
Страница 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Страница 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Страница 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Страница 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Страница 222: ...Reference Information Page 222 Quattro Ultima User s Guide...