Power
Instrument:
230V (+10%, -14%), 13A
Data system:
100-120V or 200-240V, 13A
Pumps:
230V (+10%, -14%), 13A
Environment
Ambient temperature:
15-28°C (59-82°F)
Short term variance (1.5 hours):
≤
2°C (
≤
4°F)
Overall heat dissipation
(excluding LC
and optional water chiller):
4.2kW maximum
Humidity:
Relative humidity
≤
70%
Water Cooling
Heat dissipation into the water:
200W
Exhausts
Rotary Pumps
The rotary pumps must be vented to atmosphere (external to the laboratory) via a
fume hood or industrial vent.
API Gas Exhaust
The API gas exhaust must be vented to atmosphere (external to the laboratory).
Caution:
The API gas exhaust line must not be connected to the rotary pump
exhaust line. In the event of an instrument failure, rotary pump exhaust could be
admitted into the source chamber producing severe contamination.
Hardware Specifications
Page 13
Quattro Ultima
User's Guide