The capillary can now be checked for flow by connecting the output from a
Harvard syringe pump to the other side of the union and setting the flow to
1 µl/min, using a micropipette to measure the flow. It is recommended that a
syringe with a volume of no more than 50 millilitres is used.
Thread the fused silica through the nebulising tip and screw in the nano-LC
chamber such that it is screwed in approximately half way.
Cut the fused silica using a tile cutter and adjust the nebulising tip further, such
that 1mm of fused silica protrudes from the tip.
Attach the nebulising gas tubing to the sprayer using an O ring and the special
screw.
Attach the sprayer assembly to the stage.
It may be necessary to alter the position of the thumbscrew underneath the
baseplate to attach the sprayer correctly.
Swing the stage into the interface using the stop and handle.
Operation
For tuning purposes it may be useful to infuse a known sample in 95% water using a
Harvard syringe pump.
Set the liquid flow to about 200 nl/min.
Switch on
Gas
at the MassLynx tune page.
Set the pressure of the gas on the regulator to approximately 0.5 bar (7 psi).
Ensure there are no leaks of gas at the sprayer, particularly where the PTFE
tubing is connected to it.
By viewing under magnification, the spray emanating from the capillary may be
examined and tuned by altering the nebulising tip such that a fine spray is observed.
Altering the gas slightly may also help in this tuning process.
Swing the stage back out of the source and place the cover over the sprayer
ensuring that the tubing coming from the sprayer is threaded correctly through it.
Lock the cover in place with two screws.
Swing the stage back into the source and alter the translation stage (in / out
direction) such that the capillary is approximately 5mm from the cone.
Select
Operate
and set
Capillary
to approximately 2.5kV.
An ion beam should now be present.
Nanoflow Electrospray
Page 167
Quattro Ultima
User's Guide
Содержание Quattro Ultima
Страница 2: ...Quattro Ultima User s Guide...
Страница 68: ......
Страница 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Страница 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Страница 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Страница 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Страница 222: ...Reference Information Page 222 Quattro Ultima User s Guide...