Calibration Failure
When calibration is performed manually there is no warning message to show that the
calibration has not met the set criteria. This must be judged by viewing the on-screen
calibration report and examining the matched peaks and statistics associated with the
report. There are a number of reasons for a calibration to fail:
•
No peaks. If the acquired calibration data file contains no peaks the calibration
has failed. This may be due to:
Lack of reference compound.
Wrong scans or wrong data file being used for the calibration.
No flow of solvent into the source.
Multiplier
set too low.
•
Too many consecutive peaks missed. If the number of consecutive peaks which
are not found exceeds the limit set in the Automatic Calibration Check
parameters then the calibration has failed. Peaks may be missed for the
following reasons:
The reference solution is running out causing less intense peaks to not be
detected.
Multiplier
is too low and less intense peaks are not detected.
The incorrect ionisation mode is selected. Check that the data has been
acquired with
Ion Mode
set to
APcI+
.
Intensity threshold
, set in the Calibration Parameters dialog box, is too
high. Peaks are present in the acquired calibration file but are ignored
because they are below the threshold level.
Either
Initial error
or
Peak window
, set in the Calibration Parameters
dialog box, is too small. The calibration peaks lie outside the limits set by
these parameters.
Maximum Std Deviation
(set in the Automatic Calibration Check dialog
box) has been exceeded.
The wrong reference file has been selected. Check that the correct file
(peg1000.ref in this case) is selected in the Calibrate dialog box.
Mass Calibration
Page 135
Quattro Ultima
User's Guide
Содержание Quattro Ultima
Страница 2: ...Quattro Ultima User s Guide...
Страница 68: ......
Страница 100: ...Data Acquisition Page 100 Quattro Ultima User s Guide...
Страница 117: ...Mass Calibration Page 117 Quattro Ultima User s Guide...
Страница 140: ...Mass Calibration Page 140 Quattro Ultima User s Guide...
Страница 212: ...Maintenance and Fault Finding Page 212 Quattro Ultima User s Guide...
Страница 222: ...Reference Information Page 222 Quattro Ultima User s Guide...