(2) Channel Check
Checks the operation related to the sensor input.
Press the
key to start the channel check. The result is displayed after a short time.
When an error occurs, the WGA-910A turns ON the control output “Abnormal channel.”
<Setting Range>
Bridge Volt.:
Checks that the measured value of the excitation voltage (bridge voltage) is within ±10% of
10 V or 2V.
Input Over:
Checks that the sensor output is within the input range.
Display Over:
Checks that the measured value is within ±99999.
<How to operate>
1) Press the
key on the “
Self-Check
” window.
2) Press the
key to start the channel check.
3) The result is displayed after a short time.
Only the “
Display Over”
is displaying
Error
in the right
window.