47
5-3-5. Numeric Value Registering Calibration
The numeric value registering calibration calibrates data by changing the initial value.
Execute the numeric value registering calibration after the actual load calibration and sensitivity registering calibration.
The numeric value registering calibration is used for pressure transducers for absolute pressure measurement or when
already load is applied and unable to obtain the actual load state. By registering the initial value of the transducer
written in the Test Data Sheet or registering an already measured initial value, a reference point equivalent to unload
state can be registered.
<Setting Range>
-3.2000 to +3.2000 mV/V
<How to operate>
1) Press the
key on the “Calibration (1/2)” window.
2) Press the
key on the “Calibration (2/2)” window.
3) The initial value, registered during the no-load zero, appears.
In the case of this example, the “0.1000 mV/V” appears.
4) Enter the new initial value with the numeric keyboard.
In the case of this example, as you enter “-0.5432”, the “-
0.5432mV/V” appears.
5) Press the
key.
The “
Registering ini Val.
” window blinks on the top left for
registering the new initial value.
Registering
Registering
ini. Val.
ini. val.
.
4
5
6
+/-
7
8
9
Return
C
0
1
2
3
Input the desired value with
keyboard.Push 'Register'key.
Register
New
-0.5432 mV/V
ini.Val.
Now
0.1000 mV/V
.
4
5
6
0
1
2
3
Input the desired value with
keyboard.Push 'Register'key.
Register
New
-0.5432 mV/V
+/-
7
8
9
C
Return
Numeric Reg.Cal.
ini.Val.
Now
0.1000 mV/V
Return
Measure
Calibration
Change initial value. Run
Numeric Regist Cal.
after 'Actual Load Cal.' or
'Sens.Regist.Cal.'.
2 / 2
Previous
Calibration
'Bridge Voltage','No-Load Zero
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Return
Measure
Manual Input Cal.
Next
1 / 2
TEDS Auto Cal.
TEDS Part Cal.
TEDS Ope.Setting
'TEDS Read Ope.'and 'Zero dur-
ing TEDS'are set.TEDS informa-
tions are displayed.
Calibration is automatically
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
the only selected item from
TEDS built-in sensor.
Blinks
Next
Numeric Regist Cal.
.
4
5
6
0
1
2
3
Input the desired value with
keyboard.Push 'Register'key.
Register
New
mV/V
+/-
7
8
9
C
Return
Numeric Reg.Cal.
ini.Val.
Now
0.1000 mV/V
Register
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...