41
The “
Sens. Registering”
window blinks when conducting the actual load calibration with communication
during the measurement or during the setting.
(3-b) Actual Load Cal. (Actual Load Calibration)
Actual load is applied to the sensor for the calibration. Displays the output when an actual load is applied as the
rated output. Suppose that you are using a transducer with the “Rated Capacity: 200 kN” and “Rated Output:
2.0040 mV/V.” In order to read “200.00” against a load of 200 kN that you will give to the transducer, do as
follows.
<Setting range>
Rated Output: (-3.2000 to +3.2000 mV/V; display only, excluding 0)
Rated Display: -99999 to 99999 (count; excluding 0)
<How to operate>
1) When a load is applied to the sensor, the output is displayed to the
“
New”
column of the rated output.
Check that the value is stabled.
[2.0040 mV/V] is displayed in the right figure.
2) Input a setting value of the “Rated Output” with the numeric
keyboard.
Suppose “200.00” is input, “200.00 kN” is displayed in the “
New”
column.
3) Press the
key. The “
Sens. Registering”
window on the
upper left blinks and the “Rated Output” and “Rated Display” are
registered.
4) After the registration, press the
key to move to the “
Unit
”
window.
MEMO
Blink
Sens.Regis-
Measure
Sensitivity reg-
tering
istering
Calibration is conducted with
Calibration
sensor by 'Manual Input','TEDS
Auto'or'TEDS Part'Cal.
'Smoothing','Additional Value'
Input Setting
and 'Zero Compensation'are set
The sensor output is monitored
Next
1 / 2
'Comparison Set'and 'Measure
Meas.Cond.Setting
Mode Set'are set as measure-
ment condition.
The necessary setting to dis-
Wave Display Setting
play the wave such as 'X-axis'
and 'Y-axis'are set.
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...