5-3. HOW TO CALIBRATE
5-3-1. Calibration Window
<How to operate>
1) Press the
key on the “
Measuring
” window.
2) Press the
key on the “
Setting
” window.
5-3-2. Manual Input Calibration
Conducts calibration by manual setting in order of excitation voltage, no-load zero, Sensitivity registering calibration
or in order of actual load calibration and unit.
(1) Bridge Voltage (Excitation Voltage)
Capable of setting the excitation voltage that applies to the sensor from 2 V or 10 V.
Select the higher voltage within the recommended excitation voltage of the sensor.
<Setting range>
2 V or 10 V
<How to operate>
1) Press the
key on the “
Calibration
” window.
2) Press either
or
key and press the
key to register the setting value.
3) After the registration, press the
key to move to the
“
No-load Zero
” window.
Calibration
'Bridge Voltage','No-Load Zero
,'Sens.Regist.Cal'or'Actual
Load Cal'and 'Unit'are set.
Return
Measure
Manual Input Cal.
Next
1 / 2
TEDS Auto Cal.
TEDS Part Cal.
TEDS Ope.Setting
'TEDS Read Ope.'and 'Zero dur-
ing TEDS'are set.TEDS informa-
tions are displayed.
Calibration is automatically
conducted with all reading
from TEDS built-in sensor.
Calibration is conducted with
the only selected item from
TEDS built-in sensor.
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...