13
2-1-5. Measuring Window (Number View Dual Display for [Block Peak-Bottom] and [Time Peak-Bottom])
1)
Hold
key
Manually starts a detection and holding.
2)
Switch
key
Display changes in order from “Number View Dual Display” - “Zoom In
Number View Dual Display” - “Waveform View Dual Display” by pressing the
Switch
key.
3)
[
Detecting
] lamp
[Holding]
lamp
[
Detecting
] lamp lights up in white during detection and [
Holding
] lamp lights
up in white during holding respectively.
4)
Comparator status lamp for
peak value
Based on the detecting peak value, when turning ON the comparator outputs,
the [
HI-1
] light up in red, [
OK
] lights up in white, and [
LO-1
] lights up in
yellow.
5)
Peak value
Displays a peak value with the unit.
6)
Comparator status lamp for
bottom value
Based on the detecting bottom value, when turning ON the comparator outputs,
the [
HI-1
] light up in red, [
OK]
lights up in white, and [
LO-1
] lights up in
yellow.
7)
Bottom value
Displays the bottom value with the unit.
8)
ZERO
key
Conducts digital zero to set the display of the measured value to 0 (or additional
value).
9)
Set
key
Displays the setting window.
10) Operation mode
Displays “operation mode” of the “measuring function.”
11) Measuring condition No.
Displays the current measuring condition setting No.
12) Compared value setting key
of the peak value
Displays the “HI-1” or “LO-1” window of the peak value.
13) Compared value of the peak
value
Displays the “HI-1” or “LO-1” value of the peak value.
14) Compared value setting key
of the bottom value
Displays the “HI-2” or “LO-2” window of the bottom value.
15) Compared value of the
bottom value
Displays the “HI-2” or “LO-2” value of the bottom value.
kN
-999.99
kN
Bottom
LO
OK
HI
HI-2
500.00
LO-2
-999.99
Peak
LO
OK
HI
HI-1
500.00
LO-1
Detecting
Holding
Block Peak-Bottom
Measure 1
Switch
Hold
Zero
Set
1)
2)
3)
4)
5)
6)
7)
8)
9)
10)
11)
12)
13)
14)
15)
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...