120
8-3-3. Executing
(10) [HLD]
Content:
Hold command
Format:
HLD<CR>
Parameter:
None
Return value format:
ok<CR><LF>
Description and others
Relevant to the
Hold
key operation on the touch panel.
For reference:Arbitrary point hold
HLD <CR>
Holding interval
RST <CR>
Peak hold
HLD <CR>
Detecting interval
RST <CR>
Interval-specified peak hold
HLD <CR>
Detecting interval
Holding interval
RST <CR>
Time-specified peak hold
HLD <CR>
Detecting in Holding interval
RST <CR>
(11) [RST]
Content:
Reset command (To execute the peak hold, etc.)
Format:
RST<CR>
Parameter:
None
Return value format:
ok<CR><LF>
Working properly.
(12) [ZON]
Content:
ZERO command(To execute the DIGITAL ZERO function.)
Format:
ZON<CR>
Parameter:
None
Return value format:
ok<CR><LF>
Working properly.