![KYOWA WGA-910A Series Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/kyowa/wga-910a-series/wga-910a-series_instruction-manual_2017887056.webp)
52
For replacing the sensor to the same model and same capacity one because of the sensor’s faulty, set the
Rd.
Out.
of the
TEDS Cal. Item
in the “
TEDS Part Cal.
” window to “
Execute
” and
Ex. Volt
,
Rd. Cap.
and
Unit
to “
Prohibit
” for the easy calibration.
(4) Supplemental explanations of the TEDS function
(1) Executing the TEDS calibration at the initial value.
Set the sensor in no-load state and conduct the calibration with the TEDS. The no-load zero is automatically
conducted and the display value is set to “0 (or additional value)” when the
Execute
is selected in the “
ZERO
during TEDS
” window.
Then automatically conducts the sensitivity registering calibration, updates calibration values of the excitation
voltage, rated output, rated display (including a decimal digit position), and unit.
Capable of executing the TEDS calibration by the key operation on the window, control input, when turning ON
the power and by communication command.
However, executing no-load zero to the initial value when turning ON the power is prohibited.
(2) Applied voltage
The applied voltage should be equal or less to the recommended applied voltage and higher one should be
selected.
[Example]
2 V is set for the recommended applied voltage of 3 V.
(3) Calculation of the calibration coefficient
Register the calibration coefficient that is obtained from the following expression.
Rated capacity of the sensor (y) ÷ Rated output (x)
(4) No-load zero
No-load zero of the TEDS calibration is same with that of the manual input calibration.
(5) Display value and unit
Decimal point of the rated capacity (display) and unit after conducting the TEDS calibration is automatically set
corresponding to the total number of display digits of 5.
[Example]
5 N = 5.0000 N, 500 N = 500.00 N, 50 N = 50.000 N, 5 kN = 5.0000 kN
(6) Multiple connection of the TEDS-compatible sensor
Combined rated capacity and rated output are calculated as follows when connecting the multiple TEDS-
compatible sensors (2 to 4) in parallel.
Rated Capacity (combined): Total value of the rated capacities of the sensors.
Rated Output (combined): Average value of the rated outputs of the sensors (Value obtained by dividing the
total value of sensors with the number of units.).
(7) Model, serial No., and calibrated date of the TEDS
For reading the KYOWA made sensor only.
The information may not be read when connecting the other made TEDS-compatible sensor.
MEMO
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...