![KYOWA WGA-910A Series Скачать руководство пользователя страница 127](http://html.mh-extra.com/html/kyowa/wga-910a-series/wga-910a-series_instruction-manual_2017887127.webp)
123
(21) [VRC]
Content:
To execute numeric value registering calibrations
Format:
VRC±x.xxxx<CR>
Parameter:
±x.xxxx
Initial value of transducers: -3.2000 to 3.2000 (mV/V)
Return value format:
ok<CR><LF>
E-00<CR><LF>(When an error occurs)
Description and others
The results of the command will be saved.
Reading command:
PRR011
Return value
x.xxxx<CR><LF>
Initial value of transducers:
(22) [EID]
Content:
Conduct TEDS-based automatic calibration.
Format:
EID<CR>
Parameter:
None
Return value format:
ok<CR><LF>
(when in normal)
E-13<CR><LF> (When an error occurs)
Description and others
Result of this command is overwritten.
Sets the excitation voltage and sensitivity registering calibration with the rated capacity
value, rated output value, recommended excitation voltage value, and unit of the TEDS
information.
Reading command
PRR010
Return value
±x.xxxx, ±yyyyyy, zz<CR><LF>
Unit
Rated display value
Rated output value
(23) [SSM]
Content:
Set the smoothing function.
Format:
SSMaa, b, c<CR>
Parameter:
aa, b, c
Analog filter
0: 1 Hz
1: 30 Hz
2: 300 Hz
3: None (1 kHz or more)
Minimum scale
0: 1
1: 2
2: 5
3: 10
4: 20
5: 50
6: 100
Moving average
00: None
01: 2 times
02: 4 times
03: 8 times
04: 16 times
05: 32 times
06: 64 times
07: 128 times
08 256 times
09: 512times
10: 1024 times 11: 2048 times
Return value format:
ok<CR><LF>
Description and others
Corresponding to the [/M] command.
Reading command
PRR020
Return value
aa, b, c<CR><LF>
Analog filter
Minimum scale
Moving average
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...