40
Set the rated display 100000 count or less per rated output 0.1 mV/V.
The “
Sens. Registering”
window blinks when conducting the sensitivity registering calibration with
communication during the measurement or during the setting.
(3-a) Sensitivity Regist. Cal. (Sensitivity Registering Calibration)
Inputs the rated capacity and rated output described in the Test Data Sheet for equivalent calibration with the
actual load when the actual load cannot be applied to the sensor. Suppose that you are using a transducer with the
“Rated Capacity: 200 kN” and “Rated Output: 2.0040 mV/V.” In order to read “200.00” against a load of 200 kN
that you will give to the transducer, do as follows.
<Setting range>
Rated Output: -3.2000 to +3.2000 mV/V (excluding 0)
Rated Display: -99999 to 99999 (excluding 0)
<How to operate>
1)After pressing the
key, input a setting value with a numeric
keyboards. In the case of example, if “2.0040” is input, “2.0040
mV/V” is displayed in the “
New”
column.
In addition, the
key blinks during the setting.
Press the
key. The “
Sens. Registering”
window on the
upper left blinks and the “Rated Output” is registered.
2) After pressing the
key, input a setting value with a numeric
keyboards.
In the case of example, if “200.00” is input, “200.00 kN” is displayed
in the “
New”
column.
In addition, the
key blinks during the setting.
3) Press the
key. The “
Sens. Registering”
window on the
upper left blinks and the “Rated Output” is registered.
4) After the registration, press the
key to move to the “
Unit
”
window.
Blinks
NOTE
MEMO
Sens.Regis-
Measure
Sensitivity reg-
tering
istering
Calibration is conducted with
Calibration
sensor by 'Manual Input','TEDS
Auto'or'TEDS Part'Cal.
'Smoothing','Additional Value'
Input Setting
and 'Zero Compensation'are set
The sensor output is monitored
Next
1 / 2
'Comparison Set'and 'Measure
Meas.Cond.Setting
Mode Set'are set as measure-
ment condition.
The necessary setting to dis-
Wave Display Setting
play the wave such as 'X-axis'
and 'Y-axis'are set.
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...