70
・
Short circuit the terminals between the Control I/O connector No. 35 or 36 (Input COM) and No. 20 (Hold
Command). The hold command is turned ON and the WGA-910A operates with the preset measuring
mode.
・
Short circuit the terminals between the Control I/O connector No. 35 or 36 (Input COM) and No. 21
(Reset). The reset command is turned ON and the detection and holding are released.
6-3-4. Descriptions of the operation mode
6-3-4-1. Normal
Normal measurement that always measures and displays value.
No control signal for the detection.
“Detect Time”, “Delay Time”, “Comparison Mode”, “Display Mode” cannot be set.
6-3-4-2. Peak Hold
Capable of setting the “Delay time.” However, the “Detect Time”, “Comparison Mode”, and “Display Mode” cannot be set.
(A) Behavior with control input signal
When the HOLD command is turned ON
and after the lapse of delay time, the peak
value is detected and indicated until the
HOLD command is turned OFF.
The detection of the peak value and hold
display are released when the hold command
is turned OFF or reset command is turned
ON.
(B) Behavior with key operation
When the
key is pressed the first
time and after the lapse of the delay time, the
detecting block starts and displays the peak
value until the key is pressed for the second
time.
Press the
key for the second time to
end the detecting block (reset).
MEMO
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...