60
and
keys are displayed in gray when selecting the “
Control Input
” and “
Command
” in the
“
Meas Select Signal
” window. Therefore the measuring condition No. cannot be changed.
The No. based on the ON/OFF combination of the “Measurement condition select 0 to 3” are displayed
when the selecting the “
Control Input
.”
The received command No. from the external device is displayed when selecting the “
Command
”.
6-3. MEAS. COND. SETTING (SETTING MEASURING CONDITION)
Capable of storing setting values of the comparison setting and measuring mode setting to the measuring condition file
and capable of changing the measuring condition No. by the key operation, control input, or communication command.
The number of the measuring condition files is 32 for the key operation, 16 for the control input, and 32 for the
communication command.
The setting values are stored in a file having the displaying No. on the “
Meas. Cond. Setting
” window.
6-3-1. Measuring Condition
Capable of selecting the No. of the measuring condition by the key operation on the window when selecting the
“
Key Operation
” in the “
Meas. Select Signal
” of the “
Initial Set”
window.
Capable of selecting the No. with the combination of the ON/OFF of the control input “Measurement condition
select 0 to 3” when selecting the “
Control Input
” in the “
Meas. Select Signal
.”
The No. of the measuring condition on the window also changes at the same time.
<How to operate>
For the “
Key Operation
” in the “Meas. Select Signal” of the “
Initial Set
window only.
1) Press the
key on the “
Measuring
” window.
2) Press the
key on the setting window.
3) Press the
key to change the number of the [Measure] in the
lower window.
MEMO
Содержание WGA-910A Series
Страница 112: ...108 7 OPERATION TIMING 7 1 CONTROL INPUT HOLD COMMAND RESET COMMAND ZERO COMMAND 7 1 1 Control Input ...
Страница 113: ...109 7 1 2 Block specified 7 1 3 Time specified ...
Страница 114: ...110 7 2 COMPARATOR OUTPUT 7 3 DELAY TIME BETWEEN INPUT AND OUTPUT Delay time between input RS output MAX 30 ms ...
Страница 145: ...141 11 2 DIMENSIONS Unit mm ...
Страница 150: ...146 Kyowa Electronic Instruments Co Ltd 3 5 1 Chofugaoka Chofu Tokyo 182 8520 Japan http www kyowa ei com ...