55
Decimal positions with displaying “0” all the time are not displayed based on the zero
suppression.
[Example]
Suppose the rated display: 500.00.
The setting value of the minimum scale is [0.01, 0.02, 0.05, 0.1, 0.2, 0.5, and 1].
(2) Minimum Scale (division)
Sets the minimum digit that changes the number.
However, the minimum scale not only suppresses the variation of the display value but also not to update the
display value higher than the requiring resolution.
Displays the setting values of the minimum scale according to the preset decimal position that was set in the rated
display of the calibration.
<Setting Value>
1, 2, 5, 10, 20, 50, 100 (count)
<How to operate>
1) Press the
key on the “
Smoothing
” window.
2) Press the
key to change pages until the target value is
displayed.
3) Select and press the target key and press the
key to
register the setting value.
MEMO