6
Calibration Procedures
Adjustments procedures
214
U3606A User’s and Service Guide
6
Press
to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
•
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
•
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
7
Repeat
through
for each output adjustment item shown in
.
8
Repeat
through
for the S1 (30 V/1 A) range. Press
Voltage
to select the S1 range (the S1 annunciator is illuminated).
For adjustment item “LOAd”, connect an additional 30
Ω
, 50 W load
across the front panel
(red) and
(black) output terminals. Leave
the connections from the output terminals (
and
) to the input
terminals (
V
and
LO
) intact.
N O T E
Always complete tests in the same order as shown in
O U T
S B Y
30
Ω
, 50 W load
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...