![Keysight U3606A Скачать руководство пользователя страница 281](http://html.mh-extra.com/html/keysight/u3606a/u3606a_users-and-service-manual_1961516281.webp)
Characteristics and Specifications
8
DC Power Supply Specifications
U3606A User’s and Service Guide
259
[2] For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is added. The accuracy of the duty cycle should be
calculated as:
Calculation example:
Frequency setting = 4800 Hz, Duty cycle setting = 50%
The duty cycle accuracy (for frequency setting 4800 Hz) is calculated as 50% ± 2.55%.
[3] Specification applies when the positive or negative pulse width is greater than 50
μ
s.
[4] For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is added. The accuracy of the pulse width should
be calculated as:
Calculation example:
Frequency setting = 4800 Hz, Duty cycle setting = 50%
The pulse width accuracy (for frequency setting 4800 Hz and duty cycle setting 50%) is calculated as 0.2083 ms ±10.625
μ
s.
Accuracy
0.4%
frequency
100
-----------------------------
1
–
⎝
⎠
⎛
⎞
0.1%
×
+
⎝
⎠
⎛
⎞
=
Specification
of
duty
cycle
Accuracy
Duty
cycle
setting
×
(
)
±
=
Specification
of
duty
cycle
0.4%
4800
100
------------
1
–
⎝
⎠
⎛
⎞
0.1%
×
+
⎝
⎠
⎛
⎞
50%
×
⎝
⎠
⎛
⎞
±
5.1
100
---------
50%
×
⎝
⎠
⎛
⎞
±
2.55
±
%
=
=
=
Accuracy
0.4%
frequency
100
-----------------------------
1
–
⎝
⎠
⎛
⎞
0.1%
×
+
⎝
⎠
⎛
⎞
=
Specification
of
pulse
width
Accuracy
Pulse
×
width
setting
(
)
±
=
Specification
of
pulse
width
0.4%
4800
100
------------
1
–
⎝
⎠
⎛
⎞
0.1%
×
+
⎝
⎠
⎛
⎞
1
4800
------------
×
⎝
⎠
⎛
⎞
±
5.1
100
---------
1
4800
------------
×
⎝
⎠
⎛
⎞
±
10.625
±
μ
s
=
=
=
N O T E
•
The rise and fall time is less than 25
μ
s.
•
The additional load regulation is 0.15 V/A.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...