8
Characteristics and Specifications
Digital Multimeter Specifications
248
U3606A User’s and Service Guide
Frequency specifications
Table 8-3
Frequency accuracy specifications ± (% of r % of range)
Function
Range
Frequency range
1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
Frequency
[1]
Voltage path:
100 mV to 750 V
< 2 Hz
0.18 + 0.003
0.005
< 20 Hz
0.04 + 0.003
0.005
20 Hz to 100 kHz
0.02 + 0.003
0.005
100 kHz to 300 kHz
0.02 + 0.003
0.005
Current path:
10 mA to 3 A
< 2 Hz
0.18 + 0.003
0.005
< 20 Hz
0.04 + 0.003
0.005
20 Hz to 10 kHz
0.02 + 0.003
0.005
[1] For 100 mV and 1 V ranges, the measurable frequency is up to 1 MHz at 0.5 V signal . Minimum input frequency is 1 Hz.
N O T E
All frequency counters are susceptible to errors when measuring low-voltage,
low-frequency signals. Shielding inputs from external noise pickup is critical for
minimizing measurement errors.
Table 8-4
Frequency sensitivity for voltage measurement
Input range
[1]
Minimum sensitivity (rms sine wave)
20 Hz to 100 kHz
100 kHz to 300 kHz
300 kHz to 1 MHz
100 mV
50 mV
50 mV
0.5 V
1.0 V
100 mV
120 mV
0.5 V
10 V
1 V
1.2 V
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...