![Keysight U3606A Скачать руководство пользователя страница 180](http://html.mh-extra.com/html/keysight/u3606a/u3606a_users-and-service-manual_1961516180.webp)
5
Verification and Performance Tests
Performance Verification Tests
158
U3606A User’s and Service Guide
DC current gain verification test
1
Connect the calibrator to the front panel
I
(red) and
LO
(black) input
).
2
Press
to select the DC current function. The DC annunciator will
be illuminated.
3
Select each range in the order shown below. Provide the indicated input
current. Compare the measurement results to the appropriate test limits
shown in
. (Be certain to allow for appropriate source settling
when using the Fluke 5520A.)
AC voltage verification test
1
Connect the calibrator to the front panel
V
(red) and
LO
(black) input
).
2
Press
to select the AC voltage function. The AC annunciator will
be illuminated.
3
Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare the measurement results to the
appropriate test limits shown in
. (Be certain to allow for
appropriate source settling when using the Fluke 5520A.)
Table 5-4
DC current gain verification test
Input current
Range
Error from nominal 1 year
10 mA
10 mA
±0.0065 mA
100 mA
100 mA
±0.055 mA
1 A
1 A
±0.00207 A
3 A
3 A
±0.00771 A
I
V
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...