Calibration Procedures
6
Adjustments procedures
U3606A User’s and Service Guide
217
5
Verify the constant current adjustments using the
readback accuracy” on page 168
CV output adjustment procedure — rear output terminals
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
1
Turn off the instrument and connect the rear panel
+
and
–
output terminals to the
V
(red) and
LO
(black) input terminals.
+ S + +
– – –S
30
Ω
, 50 W load
N O T E
Ensure that the rear panel sense terminals (+S and –S) are shorted to the
rear panel output (+ and –) terminals. See
for more information on how to connect the
load leads.