Verification and Performance Tests
5
Performance Verification Tests
U3606A User’s and Service Guide
167
3
Set the output voltage to the full rated value (30 V for S1 range) and
current to full scale.
4
Connect an additional electronic load across the front panel
(red)
and
(black) output terminals in parallel with a oscilloscope (see
).
5
Operate the electronic load in transient operation mode between one
half of the output full rated value and the output full rated value at a
1 Hz rate with 50% duty cycle.
6
Set the oscilloscope for DC coupling, internal synchronization, and lock
on either the positive or negative load transient.
7
Adjust the oscilloscope to display transients as shown in
Note that the pulse width (t
2
– t
1
) of the transients at 50 mV from the
base line should be no more than 450 ms for the output.
Figure 5-13
Load transient response time