Verification and Performance Tests
5
Performance Verification Tests
U3606A User’s and Service Guide
163
CV load effect (load regulation)
This test measures the change in the output voltage resulting from a
change in the output current from full load to no load or vice versa.
1
Turn off the instrument and connect a digital multimeter between the
front panel
(red) and
(black) output terminals (see
2
Turn on the instrument. Press
Voltage
to select the constant voltage
mode. Ensure that the S1 (30 V/1 A) range is selected (the S1
annunciator is illuminated). Press
Shift
>
Range
again if it is not.
3
Set the output voltage to the full rated value (30 V for S1 range) and
current to full scale.
4
Connect an additional electronic load across the front panel
(red)
and
(black) output terminals in parallel with a digital multimeter
(see
).
5
Operate the electronic load in constant resistance mode, and set its
resistance to 30
Ω
. Ensure that the U3606A is operating within the
specified limit and protection values. If not, adjust the electronic load
by increasing the resistance so that the current drops slightly until the
U3606A is operating within the specified limit and protection values.
6
Output the full rated value in the order shown below. Record the output
voltage reading on the digital multimeter.
7
Within a few seconds after
, operate the electronic load in open
mode. Record the output voltage reading on the digital multimeter.
8
Compare the difference between the measurement results to the
appropriate test limits shown in
Table 5-10
Constant voltage load effect verification test
Output voltage
Range
Error from nominal 1 year
30 V with 30
Ω
load
30 V
±0.009 V
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...