5
Verification and Performance Tests
Performance Verification Tests
168
U3606A User’s and Service Guide
CC programming and readback accuracy
This test verifies that the constant current programming and readback
accuracy are within published specifications. The accuracy of the current
monitoring resistor must be 0.1% or better.
1
Turn off the instrument and connect a 0.01
Ω
current shunt monitoring
resistor across the front panel
(red) and
(black) output terminals
and a digital multimeter across the current shunt monitoring resistor
(see
).
2
Turn on the instrument. Press
Current
to select the constant current
mode. Ensure that the S2 (8 V/3 A) range is selected (the S2
annunciator is illuminated). Press
Shift
>
Range
again if it is not.
3
Output each value in the order shown below. Divide the voltage drop
(reading on the digital multimeter) across the current monitoring
resistor by its resistance. Record and compare the measurement results
to the appropriate test limits shown in
.
CC load effect (load regulation)
This test measures the change in the output current resulting from a
change in the load from full rated output voltage to short circuit.
1
Turn off the instrument and connect a 0.01
Ω
current shunt monitoring
resistor between the front panel
(red) and
(black) output
terminals and a digital multimeter across the current shunt monitoring
resistor (see
).
2
Turn on the instrument. Press
Current
to select the constant current
mode. Ensure that the S2 (8 V/3 A) range is selected (the S2
annunciator is illuminated). Press
Shift
>
Range
again if it is not.
3
Set the output current to the full rated value (3 A for S2 range) and
voltage to full scale.
4
Connect an additional electronic load across the front panel
(red)
and
(black) output terminals (see
).
Table 5-12
Constant current programming and readback accuracy verification test
Output current
Range
Error from nominal 1 year
0 A
3 A
±0.003 A
3 A
3 A
±0.0075 A
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...