Digital Multimeter Operation
Making Measurements
28
U3606A User’s and Service Guide
Measuring AC+DC voltage
The U3606A Multimeter|DC Power Supply includes a true- rms multimeter
that returns an accurate rms reading not only for sine waves, but also
other AC signals such as square, triangle, and staircase waveforms without
any DC offset. However, you may choose to return the measured AC signal
with DC offset by using the AC+DC function.
The AC+DC voltage measurement measures AC voltage with DC offset.
1
Press
again ((until the AC+DC annunciator is shown on the
display) to make AC+DC voltage measurements.
2
Connect the red and black test leads to input terminals
V
(red) and
LO
(black) respectively as shown in
.
3
Probe the test points and read the display.
V
AC+DC annunciator