Calibration Procedures
6
Calibration Process
U3606A User’s and Service Guide
181
Calibration Process
The test equipment recommended for the calibration process and
procedure is listed in
“Recommended Test Equipment” on page 180
.
The following general procedure is the recommended method to complete
a full instrument calibration.
1
Read the
“Test Considerations” on page 145
2
Perform the verification tests to characterize the instrument (See
“Verification and Performance Tests Overview” on page 153
).
3
Unsecure the instrument for calibration. (See
instrument for calibration” on page 182
4
Perform the adjustment procedures (
5
Secure the instrument against calibration. Change the calibration code if
necessary. (See
“Changing the calibration security code” on page 185
.)
6
Note the new security code (if it has been changed) and calibration
count in the instrument maintenance records.
N O T E
Ensure that you quit the adjustment mode before powering off the
instrument.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...