Getting Started
1
Product at a Glance
U3606A User’s and Service Guide
21
Table 1-4
Output terminal connections for source functions
Function
Output terminals (+ FORCE –)
Maximum output
Constant voltage output
•
Amplitude: 0 V to 30 V
[1]
•
OCP
•
S1: 0 A to 1.1 A
•
S2: 0 A to 3.3 A
•
OC:
•
S1: 0 A to 1.05 A
•
S2: 0 A to 3.15 A
Constant current output
•
Amplitude: 0 A to 3 A
•
OVP:
•
S1: 0 V to 33 V
•
S2: 0 V to 8.8 V
•
OV:
•
S1: 0 V to 31.5 V
•
S2: 0 V to 8.4 V
Ramp output
•
Amplitude:
•
CV: 0 V to 30 V
•
CC: 0 A to 3 A
•
Number of steps: 1 to 10000 steps
Scan output
•
Amplitude:
•
CV: 0 V to 31.5 V
•
CC: 0 A to 3.15 A
•
Number of steps: 1 to 100 steps
•
Dwelling time: 1 to 99 s
Square-wave output
•
Amplitude: 0 V to 30 V
•
Frequency: 27 predefined values
•
Duty cycle: 256 steps
•
Pulse width: 256 steps
[1] Limited by range selected, S1 (30 V/1 A ) or S2 (8 V/3 A). See
“Selecting a Range” on page 97
for more information on how
to select a suitable range.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...