System Related Operation
4
Utility Menu Summary
U3606A User’s and Service Guide
123
5
Press
Shift
>
Save
to save or press
Shift
>
Exit
to exit the mode without
saving.
6
Press
Shift
>
Exit
to exit the utility menu.
Configuring the ramp signal parameters
The following procedure shows you how to configure the ramp signal
parameters.
1
To access the utility menu, press
Shift
>
Utility
.
2
The first utility menu item (Error) will be shown in the upper
secondary display.
3
Press
or
until the menu item “rAMP” is shown in the upper
secondary display.
Table 4-3
Ramp signal parameters
Ramp signal
Output
Items
Constant Voltage
Constant Current
Range
S1
S2
S1
S2
Amplitude end position
[1]
[1] Amplitude start position is fixed at 0 (V or A) by default.
0 V to 30.000 V
0 V to 8.400 V
0 A to 1.0000 A
0 A to 3.0000 A
Number of steps
1 step to 10000 steps
Dwelling time
Fastest of the instrument output capability
[2]
[2] Typically ~300 ms per step.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...