3
DC Power Supply Operation
Remote Sensing
100
U3606A User’s and Service Guide
Remote sensing connections
Remote sensing requires connecting the load leads from the rear output
terminals to the load as shown below. Observe the polarity when
connecting the sensing leads to the load.
Figure 3-3
Remote sensing connections
Figure 3-4
Local sensing connections
N O T E
The metal shorting bars should be removed from the rear output and sense
terminals for remote sensing connections. For local voltage sensing
connections, the sense leads must be connected to the output terminals.
+S
+
+
–
–
–S
LOAD
Rear output terminals
Shielded two-wire cable
+S
+
+
–
–
–S
LOAD
Rear output terminals
Short bar
Short bar