5
Verification and Performance Tests
Performance Verification Tests
166
U3606A User’s and Service Guide
6
Configure the oscilloscope as follows:
i
Set the time/div range to 5 ms/div.
ii
Set the oscilloscope to acquire every single sample at the maximum
sampling rate but retain only the minimum and maximum values in a
sampling region.
iii
Set a 20 MHz cut- off frequency for better high frequency cut- off.
iv
Enable AC coupling.
v
Enable auto- triggering.
7
Allow the oscilloscope to run for a few seconds to generate enough
measurement points.
8
Obtain the maximum peak- to- peak voltage measurement as indicated in
the oscilloscope. Divide the value by 10 to get the constant voltage
peak- to- peak noise measurement. The result should not exceed
0.03 V
pp
.
9
Disconnect the oscilloscope and connect an RMS voltmeter in its place
(see
). Do not disconnect the 50
Ω
feed- thru
termination.
10
Configure the RMS voltmeter as follows:
i
Set the high pass filter to 10 Hz.
ii
Enable AC coupling.
11
Obtain the maximum rms voltage measurement as indicated in the RMS
voltmeter. Divide the value by 10 to get the constant voltage rms noise
measurement. The result should not exceed 0.002 V
rms
.
Load transient response time
This test measures the time for the output voltage to recover to within 50
mV of nominal output voltage following a load change from full load to
half load or vice versa.
1
Turn off the instrument and connect an oscilloscope between the front
panel
(red) and
2
Turn on the instrument. Press
Voltage
to select the constant voltage
mode. Ensure that the S1 (30 V/1 A) range is selected (the S1
annunciator is illuminated). Press
Shift
>
Range
again if it is not.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...