Calibration Procedures
6
Calibration Security
U3606A User’s and Service Guide
185
Changing the calibration security code
The security code can only be changed from the remote interface after the
instrument has been unsecured.
Send the
CALibration:SECurity:CODE <new_code>
command from the
remote interface after the instrument has been unsecured.
See the
CALIbration
subsystem in the
U3606A Programmer’s Reference
for more information on the
CALibration:SECurity:CODE <new_code>
command.
Resetting the security code to the factory default
If you have forgotten the correct security code, you may follow the steps
below to change the security code back to the factory default (ATU3606A).
To unsecure the instrument without the correct security code, follow the
steps below.
&
CAL:SEC:State OFF,
ATU3606A
This command unsecures the instrument for
calibration. The default security code is
ATU3606A. The security code can only be
changed when the instrument is unsecured.
&
CAL:SEC:CODE ABC1234
This command changes the security code to
“ABC1234”.
&
CAL:SEC:State ON, ABC1234
This command secures the instrument from
accidental or unauthorized adjustments.
N O T E
If you do not have a record (or have lost the record) or the security code,
first try the factory default code — “3606”, from the front panel or
“ATU3606A”, from the remote interface — before you use the procedure
below.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...