6
Calibration Procedures
Using the Front Panel for Adjustments
190
U3606A User’s and Service Guide
Using the Front Panel for Adjustments
This section describes the process used to perform adjustments from the
front panel. Refer to the
U3606A Programmer's Reference
for remote
interface commands.
Selecting the adjustment mode
Before performing any adjustments, you must first unsecure the
instrument for calibration. See
“Unsecuring the instrument for
for more information. Once unsecured, “CALib” is
displayed on the lower secondary display to indicate that you are in
adjustment mode.
Entering adjustment values
In the adjustment procedures, to enter an input calibration value from the
front panel, use these keys to position the cursor on a digit:
Key
Description
Press
to move the cursor to the left.
Press
to move the cursor to the right.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Страница 2: ......
Страница 12: ...12 U3606A User s and Service Guide ...
Страница 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Страница 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Страница 198: ...5 Verification and Performance Tests Additional Verification Tests 176 U3606A User s and Service Guide ...
Страница 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Страница 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...